The Pureline OC8200 draws in room air through a filter. A compressor sends the filtered air to a solenoid valve, which moves it into one cylinder of Purelines two-cylinder design. This two-cylinder design ensures a constant supply of oxygen; while one side is being used, the other side is being replenished. Under pressure inside the cylinder, the molecular sieve material attracts and holds the nitrogen from the room air (21% O2). The enriched oxygen (92-95% O2) is directed to the pressure regulator, a final filter, and finally out to the anesthesia machine. Remember, while one cylinder is in use, the other is refilling with enriched oxygen ensuring you a constant, reliable supply of oxygen
Features:
Complete oxygen concentrator, frame and dual DISS male outputs Extracts medical grade oxygen (USP93%) from the air and delivers on demand- up to 8 LPM Uses the industry standard DISS male thread connection Safety back-up oxygen supply (standard E tank) with fully integrated O2 flush with industry standard 30 LPM maximum flow Ability to use your oxygen flush with a simple built in switch FDA approved, UL tested Oxygen purity and flow safety alarms Quiet 53-dBA operation Made in the USA, state-of-the-art manufacturing with ISO quality and TUV SUV safety certifications Sturdy A-frame base with compact 18 square footprint Electrical power supply:Rating: 115/230V 60Hz, Average Power: 330W(avg), Protection Class: Class II, Mains Protection: 5A
Footprint: 18 x 18 Shipping box dimensions: Concentrator 19 x 14 x 27, Frame 36 x 23 x 27 Shipping weights: Concentrator 38lbs, Frame 55lbs